Model Number N/A |
Device Problem
Fracture (1260)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/07/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The customer has indicated that the product will not be returned to zimmer biomet for investigation as its location is unknown.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that the offset reamer handles tip broke during reaming the phase of the surgery.There was no known impact or consequences to the patient.Attempts have been made and no further information has been provided.
|
|
Event Description
|
No further information at the time of this report.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.The following sections were {update/corrected}.Updated: g3; h2; h3; h6.Proposed component code: mechanical (g04)-drill.Visual examination of the provided pictures identified the reamer fractured and is missing two of the sharp edges of the device.The device is covered in bio-debris.No further evaluation can be made from the provided pictures.Part and lot identification are necessary for review of device history records, neither were provided.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|