• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL CH EXPEDIUM SPINE SYSTEM SINGLE INNER SET SCREW 5.5; FIXATION, SPINAL INTERVERTEBRAL BODY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL CH EXPEDIUM SPINE SYSTEM SINGLE INNER SET SCREW 5.5; FIXATION, SPINAL INTERVERTEBRAL BODY Back to Search Results
Catalog Number 179702000
Device Problem Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/21/2023
Event Type  malfunction  
Event Description
Device report from synthes reports an event in colombia as follows: it was reported that during an orthopedic procedure on (b)(6) 2023, the screwdrivers were oversized and did not fit the screws.The implant screw arrived with a damaged thread.This report involves one expedium spine system single inner set screw 5.5.This is report 1 of 1 for (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d2b: additional device product codes: osh, mni, mnh, kwq, kwp.D9: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.E3: reporter is a j&j employee.H3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h3, h4, h6.The photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned visual analysis of the photo revealed that single-inner setscrew was observed signs of stripped from the threaded.No other issues were observed.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed for single-inner setscrew.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.A manufacturing record evaluation (mre), was not possible because the required lot code was not provided.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPEDIUM SPINE SYSTEM SINGLE INNER SET SCREW 5.5
Type of Device
FIXATION, SPINAL INTERVERTEBRAL BODY
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INT SPINE
chemin blanc 38
le locle
Manufacturer Contact
kate karberg
chemin-blanc 38
le locle 
SZ  
3035526892
MDR Report Key17681878
MDR Text Key322609786
Report Number1526439-2023-01580
Device Sequence Number1
Product Code NKB
UDI-Device Identifier10705034064307
UDI-Public(01)10705034064307
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
K200245
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/05/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number179702000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/11/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
X25 FINAL TIGHTENER; X25 FINAL TIGHTENER
-
-