COVIDIEN EVERFLEX SELF-EXPANDING PERIPHERAL STENT WITH ENTRUST DELIVERY SYSTEM LONG; STENT, SUPERFICIAL FEMORAL ARTERY
|
Back to Search Results |
|
Catalog Number EVD35-07-120-120 |
Device Problems
Malposition of Device (2616); Activation, Positioning or Separation Problem (2906)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/17/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Physician was attempting to use an everflex entrust self-expanding stent with a 6fr sheath and a 0.014 non-medtronic (boston - thruway) guidewire during treatment of a 150mm calcified lesion in the patients left mid distal superficial femoral artery (sfa).Moderate vessel tortuosity and severe vessel calcification are reported.Lesion exhibited 90% stenosis.Artery diameter reported as 7mm.Embolic protection was not used.There was no damage noted to packaging (i.E.Shelf carton, hoop/tray).There were no issues noted when removing the device from the hoop/tray.The device was prepped per the ifu with no issues identified.The lesion was pre-dilated with a 6mm pre-dilation device.The device did not pass through a previously-deployed stent.No resistance was encountered when advancing the device.The thumbscrew/lock-pin was checked for securement prior to procedure.The lock-pin was removed before deployment.Wire movement issues are reported.Physician was unable to remove device from guidewire.The stent was stuck on wire and would not deploy.When pulling everything out as a unit (wire and stent) the stent started to deploy proximal to target.The physician completed deployment of the stent.The deployment system was still attached to wire and was removed as a unit.A replacement stent was opened and deployed at target with no issues.No patient injury reported.
|
|
Manufacturer Narrative
|
Product analysis: the device was returned with the red safety tab out of the device, the stent was deployed and there was a guidewire stuck in the device.The guidewire was measured and measured as 0.014¿.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|