It was reported that a shaft hole occurred.A 2.25 x 16mm synergy drug-eluting stent (des) was selected for treatment.However, during preparation, it was noted that a contrast was leaking from the midshaft which would not allow the stent to be prepped appropriately.The procedure was completed with a different device.There were no patient complications were reported.
|
Device evaluated by mfr: synergy xd mr us 2.25 x 16mm was returned for analysis.Visual, tactile, microscopic and functional analysis was performed on the device.No issues identified with the hypotube shaft.A visual and tactile examination of the outer and inner lumen and mid-shaft section found a cut, 1cm proximal to the port exchange.There was no sign of damage, stretching or lifting of the stent struts.No signs of movement, stent was set between the proximal and distal markerbands.Balloon cones were reviewed, and no issues were noted.Balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.Bumper tip showed no signs of distal tip damage.The device was attached to an encore inflation unit and an attempt was made to inflate the device.However, a leak was observed at the cut on the outer polymer extrusion near the port exchange.
|
It was reported that a shaft hole occurred.A 2.25 x 16mm synergy drug-eluting stent (des) was selected for treatment.However, during preparation, it was noted that a contrast was leaking from the midshaft which would not allow the stent to be prepped appropriately.The procedure was completed with a different device.There were no patient complications were reported.
|