Other, other text: b3: date of event and d4: udi number is unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
Device evaluated by manufacturer and h6.Evaluation codes: updated.D3, g1,2 email is: (b)(6).One device was received.Per visual inspection, the lever from microswitch taped on the top of the unit.No other physical damage observed.Observed the microswitch was missing the lever.The complaint was confirmed.A possible cause was the customer forced temperature check into microswitch.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.Replaced microswitch.Changed o-rings and reservoir gasket.Performed preventative maintenance (pm) and calibrated the unit.The device passed all functional and delivery tests.
|