• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OAKDALE LEVEL 1 HOTLINE BLOOD AND FLUID WARMER; WARMER, THERMAL, INFUSION FLUID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OAKDALE LEVEL 1 HOTLINE BLOOD AND FLUID WARMER; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-90
Device Problems Defective Component (2292); Failure to Deliver (2338)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Event Description
It was reported that the switch that detects the tubing was broken and would not run.Patient involvement unknown.
 
Manufacturer Narrative
Other, other text: b3: date of event and d4: udi number is unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Manufacturer Narrative
Device evaluated by manufacturer and h6.Evaluation codes: updated.D3, g1,2 email is: (b)(6).One device was received.Per visual inspection, the lever from microswitch taped on the top of the unit.No other physical damage observed.Observed the microswitch was missing the lever.The complaint was confirmed.A possible cause was the customer forced temperature check into microswitch.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.Replaced microswitch.Changed o-rings and reservoir gasket.Performed preventative maintenance (pm) and calibrated the unit.The device passed all functional and delivery tests.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE BLOOD AND FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key17686458
MDR Text Key322831776
Report Number3012307300-2023-08670
Device Sequence Number1
Product Code LGZ
UDI-Device Identifier50695085820909
UDI-Public50695085820909
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001764
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL-90
Device Catalogue NumberCON-HL-90
Was Device Available for Evaluation? No
Date Returned to Manufacturer08/29/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/15/2023
Initial Date FDA Received09/05/2023
Supplement Dates Manufacturer Received01/03/2024
Supplement Dates FDA Received01/31/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/11/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-