Model Number P1190A000016
Device Problem No Audible Alarm (1019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/07/2023
Event Type  malfunction  
Event Description
Hillrom received a report from a hillrom technician stating the nurse call had malfunction.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint # (b)(4).
 
Manufacturer Narrative
The hillrom technician found the hobs and harnesses needed to be replaced.Per the hillrom service manual the advanta¿ 2 should be subject to an effective maintenance program.This will help make sure of a long, operative life for the advanta¿ 2 bed.The preventative maintenance will help to reduce downtime due to excessive wear failures.An annual service of the bed is advised in order to maintain its characteristics and performance.Sidecom communication system: inspect and test the communication junction box.Make sure the sidecom® communication system features operate correctly.Inspect the communication cable, including the male and¿female pins in the plug.Replace as necessary.A search of the hillrom maintenance records showed hillrom performed preventative maintenance on this bed on february 2023.It is unknown if the facility performed any other preventative maintenance on this bed.The technician replaced the hobs and harnesses to resolve the reported event.Based on this information, no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANTA 2 FRAME
Type of Device
BED, MANUAL
Manufacturer (Section D)
HILL-ROM MEXICO
ave.del telefono no. 200 col. huinala
apodaca, nuevo leon 66640
MX  66640
Manufacturer Contact
brad wheeler
1069 state route 46 east
batesville, IN 47006
3128199307
MDR Report Key17686615
MDR Text Key322721635
Report Number3006697241-2023-00102
Device Sequence Number1
Product Code FNJ
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 09/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberP1190A000016
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/07/2023
Initial Date FDA Received09/05/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/10/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-