OAKDALE LEVEL 1 HOTLINE BLOOD AND FLUID WARMER; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-90
Device Problem Power Problem (3010)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Manufacturer Narrative
Other, other text: b3: date of event and d4: udi number is unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported that the device had intermittent power issues (powering on/off).Patient involvement unknown.
 
Manufacturer Narrative
Evaluation codes: updated device evaluation: one device was returned for investigation.Visual inspection noted: device received very dirty with a few customer stickers on it, front cover has multiple gouges out of the paint, line cord faded, quick connect corroded, pole clamp discolored, out dated printed circuit board (pcb) and outdated power switch.Functional testing found the device turned on for a short time and then turned off.Pump was also very loud.Root cause was attributed to a faulty pcb.Possible power switch connection failure on an outdated pcb board may have occurred from wear of use; known for causing power issues.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.No action taken due to the condition of the device.It is deemed beyond economical repair and will be scrapped.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE BLOOD AND FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key17686659
MDR Text Key322710113
Report Number3012307300-2023-08689
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001764
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/05/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL-90
Device Catalogue NumberCON-HL-90
Was Device Available for Evaluation? No
Date Returned to Manufacturer08/23/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received10/13/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/17/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-