MODERN MEDICAL EQUIP MFG LTD PLUMEPEN ELITE SURGICAL SMOKE EVAC PENCIL, 10FT TUBING,QTY20; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
|
Back to Search Results |
|
Catalog Number PLP2020 |
Device Problem
Continuous Firing (1123)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/23/2023 |
Event Type
malfunction
|
Event Description
|
The sales representative reported on behalf of the customer that the plp2020, plumepen elite surgical smoke evac pencil, 10ft tubing,qty20, was being used during an unknown procedure on (b)(6) 2023 when it was reported, ¿cautery won¿t stop working.¿.There was no report of injury, medical intervention, or hospitalization for the patient.The procedure was completed using an unknown alternate device.Further assessment information was requested to confirm the meaning of the event description and we were advised that the ¿cautery would not stop.¿.This report is being raised due to the reported malfunction with potential for injury upon reoccurrence.
|
|
Manufacturer Narrative
|
The reported device is being returned to conmed for evaluation.A supplemental and final report will be filed following the completion of the device evaluation and complaint investigation.We will continue to monitor for trends through the complaint system to assure patient safety.H3 other text : device not yet received.
|
|
Event Description
|
The sales representative reported on behalf of the customer that the plp2020, plumepen elite surgical smoke evac pencil, 10ft tubing,qty20, was being used during an unknown procedure on (b)(6) when it was reported, ¿cautery won¿t stop working.¿.There was no report of injury, medical intervention, or hospitalization for the patient.The procedure was completed using an unknown alternate device.Further assessment information was requested to confirm the meaning of the event description and we were advised that the ¿cautery would not stop.¿.This report is being raised due to the reported malfunction with potential for injury upon reoccurrence.
|
|
Manufacturer Narrative
|
The device will not be returned, and no photographic evidence was provided.Therefore, a device malfunction cannot be verified.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.This is the only complaint for this lot number and failure mode within the past two years.A two-year review of complaint history revealed there has been a total of (b)(4) complaints, regarding (b)(4) devices, for this device family and failure mode.During this same time frame (b)(4) devices have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed for this reported failure, the rate of failure would be 0.000005.We will continue to monitor for trends through the complaint system to assure patient safety.
|
|