Model Number N/A |
Device Problems
Unstable (1667); Failure to Cut (2587); Patient Device Interaction Problem (4001)
|
Patient Problems
Laceration(s) (1946); Unspecified Tissue Injury (4559); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
Injury
|
Manufacturer Narrative
|
This event is recorded with zimmer biomet under (b)(4).This medwatch is being filed as an initial report.E1 phone: (b)(6).G2 foreign: canada.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that the dermatome is skipping.Event occurred outside of surgery on an unknown date.There was no reported patient harm, or delay.Due diligence is in progress, no further information is available at this time.
|
|
Event Description
|
It was reported that the air dermatome was skipping causing harm to the patient.Harm was done as it made a full thickness laceration into subcutaneous tissue at the leading edge of the harvest unnecessarily which had to be sutured in two or three patients.Due diligence is complete, there is no further information available.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).G2 foreign: canada.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4) this medwatch is being filed to relay additional information.The following sections were updated/corrected: review of the most recent repair record determined the control bar was loose.The vespel and semi-circle bearings were replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No additional event information is available.
|
|
Search Alerts/Recalls
|