• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME Back to Search Results
Model Number N/A
Device Problems Unstable (1667); Failure to Cut (2587); Patient Device Interaction Problem (4001)
Patient Problems Laceration(s) (1946); Unspecified Tissue Injury (4559); No Clinical Signs, Symptoms or Conditions (4582)
Event Type  Injury  
Manufacturer Narrative
This event is recorded with zimmer biomet under (b)(4).This medwatch is being filed as an initial report.E1 phone: (b)(6).G2 foreign: canada.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that the dermatome is skipping.Event occurred outside of surgery on an unknown date.There was no reported patient harm, or delay.Due diligence is in progress, no further information is available at this time.
 
Event Description
It was reported that the air dermatome was skipping causing harm to the patient.Harm was done as it made a full thickness laceration into subcutaneous tissue at the leading edge of the harvest unnecessarily which had to be sutured in two or three patients.Due diligence is complete, there is no further information available.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).G2 foreign: canada.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4) this medwatch is being filed to relay additional information.The following sections were updated/corrected: review of the most recent repair record determined the control bar was loose.The vespel and semi-circle bearings were replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No additional event information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIMMER AIR DERMATOME
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key17690421
MDR Text Key322730211
Report Number0001526350-2023-01084
Device Sequence Number1
Product Code GFD
UDI-Device Identifier00889024375901
UDI-Public(01)00889024375901(11)210928(10)65239067
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 01/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/06/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00880100100
Device Lot Number65239067
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/18/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/28/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other;
Patient SexPrefer Not To Disclose
-
-