BOSTON SCIENTIFIC CORPORATION ULTRATOME XL; SNARE, FLEXIBLE Back to Search Results
Model Number M00535900
Device Problems Break (1069); Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/17/2023
Event Type  malfunction  
Manufacturer Narrative
Block h6: imdrf device code a0401 captures the reportable event of cutting wire broken.
 
Event Description
It was reported to boston scientific corporation that an ultratome xl was used during an endoscopic sphincterotomy (est) procedure performed on (b)(6) 2023.During the procedure, when the handle was pulled, the cutting wire of the ultratome xl broke.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.No further information has been obtained despite good faith efforts.Note: a photo of the complaint device outside the patient was provided by the customer and showed the cutting wire was broken and kinked.
 
Manufacturer Narrative
Block h6: imdrf device code a0401 captures the reportable event of cutting wire broken.The returned ultratome xl was analyzed, and a visual evaluation noted that the cutting wire was blackened, kinked, and broken in the proximal pierce hole, which are consistent with the findings when the device was observed under magnification and per media analysis on the provided photo.Additionally, the working length was observed in good shape.No other problems with the device were noted.The reported event of cutting wire break was confirmed.Upon analysis, it was found that the cutting wire was broken, kinked and blackened.Based on the condition of the device, the problem found could have been generated if there was contact between the device and the scope during energization or if the generator exceeded the maximum of voltage during the procedure.Also, energizing the device prior to performing sphincterotomy can compromise the cutting wires integrity and cause a premature cutting wire fatigue.Once the cutting wire breaks, any attempt to remove the device from the scope can lead to the broken section hitting the working channel of the scope.This can bend the cutting wire.It is most likely that procedural or anatomical factors encountered during the use of the device could have affected the device performance and its integrity.Based on all gathered information, the most probable root cause of this complaint is adverse event related to procedure.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.
 
Event Description
It was reported to boston scientific corporation that an ultratome xl was used during an endoscopic sphincterotomy (est) procedure performed on (b)(6) 2023.During the procedure, when the handle was pulled, the cutting wire of the ultratome xl broke.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.No further information has been obtained despite good faith efforts.Note: a photo of the complaint device outside the patient was provided by the customer and showed the cutting wire was broken and kinked.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRATOME XL
Type of Device
SNARE, FLEXIBLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key17693003
MDR Text Key322767659
Report Number3005099803-2023-04720
Device Sequence Number1
Product Code FDI
UDI-Device Identifier08714729103257
UDI-Public08714729103257
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
K930022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/06/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM00535900
Device Catalogue Number3590
Device Lot Number0029460621
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/10/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/24/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-