A lead extraction procedure commenced to remove a right atrial (ra) and a right ventricular (rv) lead due to infection.Spectranetics lead locking devices (llds) were inserted into each lead to provide traction.A spectranetics 14f glidelight laser sheath was used successfully to removed the ra lead.However, during removal of the rv lead, a small pericardial effusion was detected with no drop in blood pressure.A sternotomy was performed and a right atrial appendage (raa) perforation was discovered.The repair was successful and the patient survived the procedure.The physician stated that the ra lead tip orientation may have caused the perforation.This report captures the lld providing traction within the ra lead, which may have contributed to the perforation.There was no alleged malfunction of any spectranetics devices in use during the procedure.
|
H3): the device was discarded, thus no investigation could be completed.H6): cardiac perforation is a known risk of complication with use of the lld.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
|