Model Number CDDRA500Q |
Device Problems
Difficult to Remove (1528); Failure to Disconnect (2541)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/15/2023 |
Event Type
malfunction
|
Event Description
|
Related manufacturer reference number: 2017865-2023-42296, related manufacturer reference number: 2017865-2023-42297.It was reported that a patient presented in-clinic for a lead revision due to dislodgement of the right atrial and right ventricular leads.During the procedure, the chronic device set-screw was unable to be removed from the header.The leads were explanted on (b)(6) 2023 and the chronic device was removed.The patient was stable throughout the procedure.
|
|
Event Description
|
New information received notes the lead was difficult to remove from the device a header as a result of the reported set-screw malfunction.
|
|
Manufacturer Narrative
|
The reported field event of lead connection issue was confirmed.Septum material was observed inside the rv (df4) set screws hex cavity.This did not allow the set screw to be tightened and secured properly in the field.The issue was consistent with having occurred during the procedure.The device was above elective replacement indicator (eri) when received.Telemetry, impedance, sensing, pacing and high voltage (hv) output functions of the device were tested and found to be normal.
|
|
Search Alerts/Recalls
|