It was reported that during the procedure, the implantable cardioverter defibrillator was unable to be implanted and exhibited loss of pacing output, failure to capture, failure to sense, and an impedance problem.The device was not used and another device was used to complete the procedure.No adverse patient consequences were reported.
|
The reported event of a sensing, impedance, and pacing anomalies was confirmed.Testing indicates the presence of an open circuit, which is consistent with the reported anomalies.Further testing could not be performed due to the device inadvertently entering backup mode from handling and unable to be recovered.The cause of the open circuit could not be determined.
|