• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT VR; No Match

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT VR; No Match Back to Search Results
Model Number CDVRA500Q
Device Problem Migration (4003)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/16/2023
Event Type  Injury  
Event Description
Related manufacturer reference numbers: 2017865-2023-42320.It was reported by the patient that the right ventricular lead exhibited inappropriate shock.Upon review of transmission, it was found that the right ventricular (rv) lead exhibited inappropriate shock due to t-wave over-sensing.It was also found that the rv lead exhibited r-wave amplitude variation, low lead impedance, and high capture threshold.It was believed that the patient had twiddler's syndrome as the implantable cardioverter defibrillator (icd) was found to be migrated.The rv lead was explanted and replaced, and the icd was repositioned during the same procedure on (b)(6) 2023.The patient was discharged home in good condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT VR
Type of Device
No Match
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17695592
MDR Text Key322799342
Report Number2017865-2023-42321
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067031990
UDI-Public05415067031990
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial
Report Date 09/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCDVRA500Q
Device Lot NumberP000174146
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received09/06/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/28/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age64 YR
Patient SexFemale
-
-