Model Number CD3369-40C
Device Problems Loose or Intermittent Connection (1371); Over-Sensing (1438)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/24/2023
Event Type  malfunction  
Event Description
Related manufacturer report #:2017865-2023-42333 it was reported that non sustained right ventricular oversensing due to noise signals causing brief pacing inhibition was observed on the right ventricular (rv) lead.The noise was noted to occur subsequent to a recent generator change and was not present on the original generator.A possible fracture of the rv lead or set screw issue was suspected.The noise could not be programmed around.A revision was discussed.There were no reported patient consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA MP ICD
Type of Device
No Match
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17695680
MDR Text Key322830026
Report Number2017865-2023-42334
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734508346
UDI-Public05414734508346
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 09/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCD3369-40C
Device Lot NumberA000138807
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/18/2023
Initial Date FDA Received09/06/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/26/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
QUARTET LV LEAD.; TENDRIL RA LEAD.
Patient Age81 YR
Patient SexFemale
-
-