• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. ECHO POR FEM RED LAT NC 11X135; PROTHESIS, HIPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. ECHO POR FEM RED LAT NC 11X135; PROTHESIS, HIPS Back to Search Results
Model Number N/A
Device Problem Delivered as Unsterile Product (1421)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/15/2023
Event Type  malfunction  
Manufacturer Narrative
(b)(4).G2: foreign: new zealand.Customer has indicated that the product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that after a biometric stem was opened and implanted, the team noticed two (2) oral swabs inside the same container as the stem.There was no reported impact to the patient.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.Proposed component code: swab.Visual examination of the provided pictures identified the swabs within the package as reported.A review of the device history records identified no related deviations or anomalies during manufacturing.A review of complaint history found no additional related issues for this item and the reported part and lot combination.Medical records were not provided.The root cause of the reported issue is attributed to a manufacturing issue.The event is confirmed via the provided photo and chs.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No additional event information to report at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECHO POR FEM RED LAT NC 11X135
Type of Device
PROTHESIS, HIPS
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key17698886
MDR Text Key322837802
Report Number0001825034-2023-02094
Device Sequence Number1
Product Code KWA
UDI-Device Identifier00880304462878
UDI-Public(01)00880304462878(17)250511(10)534040
Combination Product (y/n)N
Reporter Country CodeNZ
PMA/PMN Number
K070274
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/07/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberN/A
Device Catalogue Number192511
Device Lot Number534040
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/10/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/11/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
-
-