ABBOTT QUARTET; NO MATCH Back to Search Results
Model Number 1458Q/86
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/24/2023
Event Type  malfunction  
Event Description
It was reported that the left ventricular (lv) lead was damaged during the initial implant procedure.The lv lead was explanted and replaced to resolve the event.The patient was in stable condition.
 
Manufacturer Narrative
The reported event of lead damage was confirmed.As received, a complete lead was returned in one piece.Visual inspection found the connector pin and molded cap were pulled out of the connector assembly, stretching the inner coil consistent with damage cause during the procedure.Evidence of solvent bonding were observed around the molded cap and inside the connector assembly.The cause of the reported event was consistent with damages occurring while applying forces between the lead body and pin resulting in the connector pin being pulled out and becoming separated from the connector assembly, consistent with procedural damage.A review of the device history record (dhr) confirmed that no issues were identified related to this reported event,.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
NO MATCH
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17699442
MDR Text Key322847958
Report Number2017865-2023-42392
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)Y
Reporter Country CodeFR
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/07/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1458Q/86
Device Lot NumberS000090397
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/03/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/03/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-