Model Number P1190A000074 |
Device Problem
Unintended Movement (3026)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/09/2023 |
Event Type
malfunction
|
Event Description
|
Hillrom received a report from a hillrom technician stating the bed was going up and down by itself when siderail was up.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint # (b)(4).
|
|
Manufacturer Narrative
|
The hillrom technician found the membrane needed to be replaced.Per the hillrom service manual the advanta¿ 2 should be subject to an effective maintenance program.Test each of the buttons to check that they activate the correct function and that they do not work intermittently by pressing each button for several seconds.Each movement must be continuous.Replace the pendant if necessary.A search of the hillrom maintenance records did not show hillrom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The technician replaced the membrane to resolve the reported event.Based on this information, no further action is required.
|
|