It was reported by the healthcare professional in china that during a shoulder scope procedure on (b)(6) 2023, it was observed that the anchor on the lupine loop rapide anchor w/orthocord ds device was broken off upon opening its package.During in-house engineering evaluation, it was determined that the the anchor, shaft as well as the suture were partially impregnated of foreign matter, presumably biological matter.Another like device was used to complete the procedure.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: udi: (b)(4).Investigation summary: the device was returned to mitek for evaluation.Mitek then conducted visual inspection of device received.Upon visual inspection of the device, it was observed that the device was returned in used condition, the anchor does not show broken areas.The anchor, shaft as well as the suture were partially impregnated of foreign matter, presumably biological matter the suture was completely out of the device, the inserter tip and the shaft were in normal shape.A manufacturing record evaluation was performed for the finished device lot number: 8l94753, and no non-conformances were identified.As part of mitek¿s quality process all devices are manufactured, inspected, and released to approved specifications.Based on the condition in which the device was returned, and since no broken parts in the anchor were identified, this complaint cannot be confirmed and a root cause for the issue experienced by the customer cannot be established.At this point in time, no corrective action is required, and no further action is warranted.However, in depuy synthes mitek, additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.
|