It was reported that a patient presented with bvvi mode noted on the patient's implantable cardioverter defibrillator.While corrective programming changes were being made, the patient reported lightheadedness, fatigue, and heart palpitations with discomfort.Following completion of the programming changes, the patient went into ventricular fibrillation, resulting in syncope and loss of consciousness.Anti tachycardia pacing was delivered but was not successful in converting the rhythm.The device reverted to bvvi mode and the patient's symptoms were resolved.It was suspected that the patient had become bradycardic or the device was over-sensing noise, so the device was explanted and replaced on (b)(6) 2023.No further adverse consequences were reported and the patient was stable.
|
The reported event of sensing noise was confirmed.External noise was confirmed and resulted in oversensing.The device appears to behave as programmed.The reported event of high voltage output issue was not confirmed.The reported event of backup operation was confirmed.Analysis of device image found the device went into backup mode due to memory error.After the device was restored from backup mode, functional testing was performed.Telemetry, impedance, sensing, pacing and high voltage (hv) output functions of the device were tested and found to be normal.The backup operation could not be reproduced.The cause of the reported event could not be determined.
|