Model Number CD3371-40QC
Device Problem Over-Sensing (1438)
Patient Problem Dizziness (2194)
Event Date 08/23/2023
Event Type  malfunction  
Event Description
It was reported that the patient presented for a routine check in clinic.The patient reported that he gets dizzy when lifting items above his head and demonstrated the manoeuvre.Oversensing on the implantable cardioverter defibrillator (icd) and right ventricular (rv) lead due to noise was observed, but the patient was asymptomatic to the manoeuver.Non sustained oversensing due to noise had also been observed (b)(6) 2023 but the patient did not recall having any symptoms.Atrial lead noise resulting in auto mode switching was also observed.Programming changes to right ventricular sensing was made.The patient was stable before, during and after the programming changes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA MP ICD
Type of Device
No Match
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17705821
MDR Text Key323161882
Report Number2017865-2023-42500
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 09/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2022
Device Model NumberCD3371-40QC
Device Lot NumberA000096413
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/23/2023
Initial Date FDA Received09/07/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/27/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-