MAUDE Adverse Event Report: STERIS MEXICO, S. DE R.L. DE C.V. V-PRO MAX STERILIZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STERIS MEXICO, S. DE R.L. DE C.V. V-PRO MAX STERILIZER Back to Search Results
Device Problem Use of Device Problem (1670)
Patient Problem Burn(s) (1757)
Event Date 08/09/2023
Event Type  malfunction  
Manufacturer Narrative
A steris service technician arrived onsite following the reported event to inspect the v-pro max sterilizer and found the unit to be operating properly.No issues were noted with the function or operation of the unit and the sterilizer was returned to service.The technician was informed that the employees were not wearing proper ppe, specifically gloves, while operating the v-pro max sterilizer.The v-pro max sterilizer operator manual (1-2) states, "danger - chemical injury hazard: any visible liquids in the chamber or on the load must be treated as concentrated hydrogen peroxide.Observe all hydrogen peroxide handling precautions.When handling hydrogen peroxide, wear appropriate personal protective equipment." the v-pro max operator manual (1-2) states, "ansi/aami st58, 2013, recommends using chemical-resistant gloves when using the sterilization unit." additionally, user facility personnel should ensure all instruments are properly dry prior to placement in the v-pro max sterilizer.The v-pro max operator manual, (a-1) states, "dry all items thoroughly.Ensure all moisture is removed from all internal parts (including lumens).If not, residual hydrogen peroxide may remain at cycle completion and/or a cycle abort occurs.Only dry items are to be placed in sterilization unit." the technician counseled user facility personnel on the importance of wearing proper ppe, specifically gloves, while operating their v-pro max sterilizer and properly drying instruments prior to processing.No additional issues have been reported.
 
Event Description
The user facility reported that two employees obtained a burn to their hands after handling an instrument pack that was processed in their v-pro max sterilizer.The employees rinsed their hands with water and sought medical treatment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V-PRO MAX STERILIZER
Type of Device
STERILIZER
Manufacturer (Section D)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX  67190
Manufacturer (Section G)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX   67190
Manufacturer Contact
daniel davy
5960 heisley road
mentor, OH 44060
MDR Report Key17705973
MDR Text Key323058221
Report Number3005899764-2023-00044
Device Sequence Number1
Product Code MLR
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/07/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/09/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/12/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-