LIVANOVA DEUTSCHLAND S5 DOUBLE HEAD PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-85-00
Device Problems Pumping Stopped (1503); No Apparent Adverse Event (3189)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/10/2023
Event Type  malfunction  
Event Description
Livanova deutschland received a report that a s5 double head pump gave an error message during procedure and administration of cardioplegia solution was not possible.The issue was improved once by turning off and on the double pump, however the error message was displayed again.There was no patient injury.
 
Manufacturer Narrative
A.1.-a.5.There was no patient involvement.H10: livanova deutschland manufactures the s5 double head pump.The incident occurred in (b)(6).Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Manufacturer Narrative
H10: through follow-up communication livanova learned that following some control taken on the pump by the customer, it stopped.However, no specific error message was shown on the s5 system control panel; only a red symbol related to cardioplegia.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Manufacturer Narrative
H10: through follow-up communication livanova learned the correct device serial number which was used during the surgery.Model/serial numbers have been added to the dedicated d.4 section.The serial read-out of the pump (real time device parameters and setting recording file) and pictures provided by the customer were analyzed indicating that the pump correctly stopped due to the activation of a monitoring function, according to device instructions for use.No technical error message was stored in the serial read out in the date of the event and the pictures provided show the icon of a pump stop associated to a monitoring function activated.Therefore, the event has been re-assessed as not reportable since the roller pump correctly stopped due to the activation of a monitoring function and no malfunction occurred.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 DOUBLE HEAD PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80939
GM   80939
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key17709179
MDR Text Key322938261
Report Number9611109-2023-00427
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup,Followup
Report Date 09/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-85-00
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/10/2023
Initial Date FDA Received09/08/2023
Supplement Dates Manufacturer Received08/28/2023
11/10/2023
Supplement Dates FDA Received09/27/2023
12/06/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured05/29/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-