Catalog Number 22438-19 |
Device Problem
Difficult to Remove (1528)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/09/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that the procedure was to treat the carotid artery with heavy calcification and heavy tortuosity.The emboshield nav6 embolic protection filter was prepped, advanced and deployed.When the physician wanted to remove the sheath [delivery catheter], it was stuck and could not be removed.Therefore, the whole emboshield nav6 eps was removed without issue.A new emboshield nav6 eps was used to complete the procedure.There were no adverse patient effects and there was no clinically significant delay in the procedure.No additional information was provided.
|
|
Manufacturer Narrative
|
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
|
|
Manufacturer Narrative
|
Visual analysis was performed on the returned product.The reported difficulty to remove was unable to be confirmed due to the condition of the returned device.However, the guidewire exit port was noted to be torn, stretched and jagged which likely contributed to the difficulty removing the delivery catheter.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar complaints reported from this lot.The investigation determined that the reported difficulty was due to procedural related circumstances.It is likely that during use the clearance between the inner diameter of the delivery catheter and outer diameter of the barewire was reduced, likely due to anatomical conditions, resulting in difficulty removing the delivery catheter due to the barewire tearing into the guidewire exit port of the delivery catheter during the attempt to remove.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
|
Event Description
|
It was reported that the procedure was to treat the carotid artery with heavy calcification and heavy tortuosity.The emboshield nav6 embolic protection filter was prepped, advanced and deployed.When the physician wanted to remove the sheath [delivery catheter], it was stuck and could not be removed.Therefore, the whole emboshield nav6 eps was removed without issue.A new emboshield nav6 eps was used to complete the procedure.There were no adverse patient effects and there was no clinically significant delay in the procedure.No additional information was provided.
|
|
Search Alerts/Recalls
|
|