• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR EMBOSHIELD NAV6 EMBOLIC PROTECTION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR EMBOSHIELD NAV6 EMBOLIC PROTECTION SYSTEM Back to Search Results
Catalog Number 22438-19
Device Problem Difficult to Remove (1528)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/09/2023
Event Type  malfunction  
Event Description
It was reported that the procedure was to treat the carotid artery with heavy calcification and heavy tortuosity.The emboshield nav6 embolic protection filter was prepped, advanced and deployed.When the physician wanted to remove the sheath [delivery catheter], it was stuck and could not be removed.Therefore, the whole emboshield nav6 eps was removed without issue.A new emboshield nav6 eps was used to complete the procedure.There were no adverse patient effects and there was no clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
 
Manufacturer Narrative
Visual analysis was performed on the returned product.The reported difficulty to remove was unable to be confirmed due to the condition of the returned device.However, the guidewire exit port was noted to be torn, stretched and jagged which likely contributed to the difficulty removing the delivery catheter.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar complaints reported from this lot.The investigation determined that the reported difficulty was due to procedural related circumstances.It is likely that during use the clearance between the inner diameter of the delivery catheter and outer diameter of the barewire was reduced, likely due to anatomical conditions, resulting in difficulty removing the delivery catheter due to the barewire tearing into the guidewire exit port of the delivery catheter during the attempt to remove.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event Description
It was reported that the procedure was to treat the carotid artery with heavy calcification and heavy tortuosity.The emboshield nav6 embolic protection filter was prepped, advanced and deployed.When the physician wanted to remove the sheath [delivery catheter], it was stuck and could not be removed.Therefore, the whole emboshield nav6 eps was removed without issue.A new emboshield nav6 eps was used to complete the procedure.There were no adverse patient effects and there was no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMBOSHIELD NAV6 EMBOLIC PROTECTION SYSTEM
Type of Device
EMBOLIC PROTECTION SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key17711772
MDR Text Key322972878
Report Number2024168-2023-09779
Device Sequence Number1
Product Code NTE
UDI-Device Identifier08717648137754
UDI-Public08717648137754
Combination Product (y/n)N
Reporter Country CodeMX
PMA/PMN Number
K141678
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/08/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number22438-19
Device Lot Number3032962
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/06/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/29/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
EMBOSHIELD NAV 6 DELIVERY CATHETER.
Patient Age48 YR
Patient SexFemale
Patient Weight78 KG
Patient EthnicityHispanic
-
-