Model Number HL-90
Device Problems Defective Alarm (1014); No Audible Alarm (1019)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Manufacturer Narrative
Other text: b3: date of event, d4:udi section is unknown.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported that the overtemp alarm on the device was not working.Patient involvement is unknown.
 
Manufacturer Narrative
Other text: h3.Device evaluated by manufacturer and h6.Evaluation codes: updated.One device was returned.Per visual inspection, the front cover had nicks and scratches, pole clamp discolored, line cord discolored, quick connect corroded, enclosure with nicks and scratches and very dirty.Per functional testing, the heater was very loud, and alarms did not go off for over temperature, add water, disposable.The light emitting diodes (led)s lit up but there were no alarms.The complaint was confirmed.The root cause was a faulty printed circuit board (pcb) speaker with no audible alarm sound.It was unknown what caused it to become faulty.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.No action taken due to the condition of the device.It was deemed beyond economical repair and was scrapped.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key17712315
MDR Text Key322979583
Report Number3012307300-2023-08769
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001764
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL-90
Device Catalogue NumberCON-HL-90
Was Device Available for Evaluation? No
Date Returned to Manufacturer08/29/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/18/2023
Initial Date FDA Received09/08/2023
Supplement Dates Manufacturer Received10/13/2023
Supplement Dates FDA Received11/03/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/30/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No