MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-3023-53
Device Problem Premature Discharge of Battery (1057)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/29/2023
Event Type  malfunction  
Event Description
It was reported during a routine check the cs300 intra-aortic balloon pump (iabp) unit battery failed after 20-30 minutes.There was no patient involvement reported.
 
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Manufacturer Narrative
A getinge field service engineer (fse) performed battery run test and observed cs300 batteries failed @58 minutes.Replaced defective batteries and corrected customer failure.Unit passed all functional and safety tests per factory specifications, returned to customer.The defective components were received for further investigation.Please refer to the root cause evaluation field for details.The failure analysis and testing dept.Received (2) part number 0146-00-0039 with a reported unit failure of a failed battery runtime test at 52 minutes.The fat performed a visual inspection and found the part to be in good condition.The fat installed the parts into cs300 test fixture serial number si206230b5 and tested the part to factory specifications per procedure number 0002-07-d016 revision e and the cs300 service manual part number 0070-00-0689 rev w.The battery runtime test failed at 54 minutes.Fat was able to verify the reported issue.Retaining the part in the failure analysis and testing department per procedure number 0002-07-d008 rev.Ap.The non-conformances with the returned components were confirmed.However, the root cause or the most probable root cause is impossible to be defined.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key17714127
MDR Text Key323000450
Report Number2249723-2023-04003
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107882
UDI-Public10607567107882
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 04/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/08/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-3023-53
Device Catalogue Number0998-00-3023-53
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/25/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/07/2010
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-