TOSOH HI-TEC, INC. AIA-900; FLUOROMETER, FOR CLINICAL USE Back to Search Results
Model Number AIA-900
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/17/2023
Event Type  malfunction  
Event Description
A customer reported error messages ¿2300 s.Loader step feed failure and 5100 assay operation aborted!¿ on the aia-900 analyzer.The customer cleaned the sample loading area, performed all-set-home tasks and rebooted the analyzer, but the errors persisted.The analyzer is down.A field service engineer (fse) was dispatched to address the reported event which resulted in a delayed reporting of patient samples for alpha-fetoprotein (afp).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Manufacturer Narrative
A field service engineer (fse) was at the customer¿s site to address the reported event.Fse confirmed the complaint.While troubleshooting, fse found the loader was out of alignment when checking the sensor flag position and suspects it was possibly due to a sample rack that crashed.The fse realigned the sample loader sensor flag for x1 axis and performed sampler loader rotation on system mainte program function with no errors.Fse repaired and validated the analyzer by successfully performing the daily check, and quality control run without error and within acceptable range.No further action required by field service.The aia-900 analyzer is functioning as expected.A 13-month complaint history review and service history review for similar complaints was performed for serial number (b)(6) from 17jul2022 through aware date (b)(6) 2023.There were five similar complaints identified during the search period including this case for error message 2300 s.Loader step feed failure, however, there was none for error 5100 assay operation aborted.The aia-900 operator's manual under section12: error messages states the following: (2300) s.Loader step feed failure cause: the pitch sensor s071 failed to go on after the step feed.Action: check to see if there is any impediment to the sample rack feed.If the trouble reoccurs, contact the tosoh local representatives.Check s071 and the step feed mechanism.(5100) assay operation aborted cause: measurement was interrupted due to the occurrence of a phenomenon that prevented the continuation of measurement.Action: check the abort cause.The most probable cause was due to misalignment of the sample loader sensor flag for x1 axis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIA-900
Type of Device
FLUOROMETER, FOR CLINICAL USE
Manufacturer (Section D)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA  7460042
Manufacturer (Section G)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA   7460042
Manufacturer Contact
jorge porras
shiba-koen first building
3-8-2 shiba
minato-ku 10586-23
JA   1058623
6506368314
MDR Report Key17717044
MDR Text Key323102673
Report Number3004529019-2023-00416
Device Sequence Number1
Product Code KHO
UDI-Device Identifier04560189283992
UDI-Public04560189283992
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K971103
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAIA-900
Device Catalogue Number022930
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/17/2023
Initial Date FDA Received09/10/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/01/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-