• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. 32" (81 CM) APPX 6.3 ML, 15 DROP ADMIN SET W/0.2 MICRON FILTER, ROTATING LUER W/; STOPCOCK, I.V. SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. 32" (81 CM) APPX 6.3 ML, 15 DROP ADMIN SET W/0.2 MICRON FILTER, ROTATING LUER W/; STOPCOCK, I.V. SET Back to Search Results
Catalog Number B30183
Device Problems Fluid/Blood Leak (1250); Air/Gas in Device (4062)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/10/2023
Event Type  malfunction  
Event Description
The event involved a 32" (81 cm) appx 6.3 ml, 15 drop admin set w/0.2 micron filter, rotating luer w/filter cap, bag hanger where an air in line towards end of infusion was reported.The event occurred at 1100 involving etoposide.There was patient involvement and unknown patient harm reported.
 
Manufacturer Narrative
One used sample list #b30183 connected with spiros, a used bd phaseal and a fresenius kabi 0.9% sodium chloride injection usp freeflex 500 ml intravenous bag with durvalumab were returned for evaluation.The inlet and outlet filters appeared to be wetted out.No additional damage or anomaly was observed.The set was primed and purged as per procedure and no leaks along the device, on the bd phaseal connection neither air in line were observed.No air in line were identified after the test.However, complaint of leaks can be confirmed based on the inlet and outlet filters that appeared to be wetted out, nevertheless during the test, leaks couldn¿t be replicated.The probable cause is typical due to a temporary or complete loss of hydrophobic properties of the filter vent material due to an infusate interaction during use.The lot history was reviewed, no nonconformities were identified that may have contributed to the reported complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
32" (81 CM) APPX 6.3 ML, 15 DROP ADMIN SET W/0.2 MICRON FILTER, ROTATING LUER W/
Type of Device
STOPCOCK, I.V. SET
Manufacturer (Section D)
ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V.
avenida cuarzo no. 250
ensenada, b.cfa. 22790
MX  22790
Manufacturer Contact
reed covert
600 n. field dr.
lake forest, IL 60045
2247062300
MDR Report Key17719345
MDR Text Key323065471
Report Number9617594-2023-00677
Device Sequence Number1
Product Code FMG
UDI-Device Identifier00887709055445
UDI-Public(01)00887709055445(17)270901(10)9930806
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K964435
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/11/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberB30183
Device Lot Number9930806
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/31/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/14/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/01/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
0.9% NACL INJECTION USP FREEFLEX 500ML IV BAG, FK; DURVALUMAB, MFR UNK; ETOPOSIDE, MFR UNK; PHASEAL, BD MFR; SPIROS, MFR UNK
-
-