ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. 32" (81 CM) APPX 6.3 ML, 15 DROP ADMIN SET W/0.2 MICRON FILTER, ROTATING LUER W/; STOPCOCK, I.V. SET
|
Back to Search Results |
|
Catalog Number B30183 |
Device Problems
Fluid/Blood Leak (1250); Air/Gas in Device (4062)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/10/2023 |
Event Type
malfunction
|
Event Description
|
The event involved a 32" (81 cm) appx 6.3 ml, 15 drop admin set w/0.2 micron filter, rotating luer w/filter cap, bag hanger where an air in line towards end of infusion was reported.The event occurred at 1100 involving etoposide.There was patient involvement and unknown patient harm reported.
|
|
Manufacturer Narrative
|
One used sample list #b30183 connected with spiros, a used bd phaseal and a fresenius kabi 0.9% sodium chloride injection usp freeflex 500 ml intravenous bag with durvalumab were returned for evaluation.The inlet and outlet filters appeared to be wetted out.No additional damage or anomaly was observed.The set was primed and purged as per procedure and no leaks along the device, on the bd phaseal connection neither air in line were observed.No air in line were identified after the test.However, complaint of leaks can be confirmed based on the inlet and outlet filters that appeared to be wetted out, nevertheless during the test, leaks couldn¿t be replicated.The probable cause is typical due to a temporary or complete loss of hydrophobic properties of the filter vent material due to an infusate interaction during use.The lot history was reviewed, no nonconformities were identified that may have contributed to the reported complaint.
|
|