COVIDIEN EVERFLEX SELF-EXPANDING PERIPHERAL STENT WITH ENTRUST DELIVERY SYSTEM STANDARD; STENT, SUPERFICIAL FEMORAL ARTERY
|
Back to Search Results |
|
Catalog Number EVX35-08-060-120 |
Device Problems
Malposition of Device (2616); Difficult to Advance (2920)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/21/2023 |
Event Type
malfunction
|
Event Description
|
A physician was attempting to implant an everflex entrust for treatment of arteriovenous fistula stenosis.It was reported there was poor guidewire movement during the procedure and the everflex stent released in vivo in a unintended location with the red lock on, it was stuck to the guide and the guide was removed and released.There was no deformation.Stent was not implanted.No patient injury reported.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Product analysis the device was returned with the red safety lock pin in place, the strain relief was confirmed as 60mm the device was returned with approx.35mm of the stent exposed, an inspection of the device found a kink approximately 13.5cm from the strain relief, the deployment wheel would not rotate with the locking tab in situ.An in-house guidewire was advanced along the length of the device but met resistance at the site of kink the red safety lock pin was removed, and the deployment wheel was moved in the direction of the arrow, it was advanced until the stent was released.The stent was examined and confirmed as 60mm.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|