It was reported during implant procedure that the left ventricular lead exhibited loss of capture and had dislodged.Attempt to advance a guidewire over the lead was unsuccessful.A stylet was then used to advance the lead, but in attempting to remove said stylet it became stuck in the lead.Upon stylet removal, the inner conductor came out with the stylet.The broken lead was successfully exchanged.The patient was stable and asymptomatic.
|
The reported events were lead dislodgement, failure to capture, difficulty inserting a stylet, stylet could not be removed and lead damage.The reported events of difficulty inserting a stylet, stylet could not be removed, and lead damage were confirmed but the reported event of failure to capture was not confirmed.As received, a complete lead with stylet stuck was returned in one piece.The ptfe coating of the stuck stylet was stripped and was found bunched up with the inner coil distal to the connector pin.The connector pin with the cap and crimp sleeve were found pulled out of the molded connector consistent with damage due to excessive forces applied while attempting to remove the stylet from the lead during the procedure.The cause of the reported events of difficulty inserting a stylet, stylet could not be removed, and lead damage was isolated to the bunching of the stylet ptfe coating inside the inner coil at the connector region that prevented the removal of the stylet and excessive forces resulted in the connector pin to be pulled out through the connector assembly.Electrical testing did not find any indication of conductor fractures or internal shorts.The s-curve height was measured within specification.
|