Model Number 4FC12 |
Device Problems
Material Twisted/Bent (2981); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Cardiac Perforation (2513); Pericardial Effusion (3271)
|
Event Date 08/24/2023 |
Event Type
Injury
|
Event Description
|
It was reported that during a cryo ablation procedure, after isolation of the left upper pulmonary vein the balloon catheter was positioned in the left inferior pulmonary vein (lipv).The mapping catheter, sheath, and balloon catheter were being manipulated.The mapping catheter was pulled in and out of the balloon catheter while the balloon catheter was resting against the atrial wall.The mapping catheter perforated through the atrium into the epicardial space.Atrial myocardial perforation with pericardial effusion was noted.It is unknown if an intervention was required.The case was aborted.The patient was not under general anesthesia.The patient's hospitalization was extended as a result of this event.No further patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Product event summary: the 4fc12 sheath with lot number 0011563070 was returned an analyzed.Visual inspection prior to disassembly and functional testing was performed on the shaft, handle.The inspection identified a shaft kink/twist at approximately two inches from the tip.The dilator was not returned.The steering mechanism and deflection test were performed.The shaft is bending as initially intended and is bending in the plane.There is no difficulty, no friction, or any noise in the steering mechanism.In conclusion, the reported perforation and pericardial effusion occurred during the procedure and the sheath failed the returned product inspection due to a kink/twist on the shaft.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|