Catalog Number 2C4705K
Device Problem No Flow (2991)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/17/2023
Event Type  malfunction  
Event Description
It was reported that a small volume folfusor did not infuse.The device line did not have any kinks and the patient's peripherally inserted central catheter (picc) line was patent.The issue occurred during patient infusion.The device was filled with vancomycin (viatris) 1000mg in sodium chloride 0.9%.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
The device was connected to the patient for 24 hours prior to being disconnected.The actual device was received for evaluation.The unit did not contain fluid in the bladder because the luer was not closed with a cap, which allowed fluid from the bladder to empty inside a bag that contained the unit.A visual inspection did not identify any abnormalities that could have contributed to the reported condition.A functional flow rate test was performed and the flow rate of the device was found to be within specification.The reported condition was not verified.The device was determined to be conforming product.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOLFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
building 3
irvine CA 92614
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key17728038
MDR Text Key323152821
Report Number1416980-2023-04592
Device Sequence Number1
Product Code MEB
UDI-Device Identifier00085412081588
UDI-Public(01)00085412081588
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date 10/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C4705K
Device Lot Number22C008
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/18/2023
Initial Date FDA Received09/12/2023
Supplement Dates Manufacturer Received09/27/2023
Supplement Dates FDA Received10/16/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/03/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SODIUM CHLORIDE 0.9%; VANCOMYCIN (VIATRIS)
-
-