• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 PUREWICK FEMALE EXTERNAL CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 PUREWICK FEMALE EXTERNAL CATHETER Back to Search Results
Catalog Number PWFX30
Device Problem Decrease in Suction (1146)
Patient Problems Bacterial Infection (1735); Sepsis (2067); Urinary Tract Infection (2120)
Event Date 08/15/2023
Event Type  Injury  
Event Description
It was reported that the patient was no longer using because the wick was backing up and not suctioning properly and caused the patient to develop multiple bacterial infections.It was unknown what medical intervention was provided for bacterial infections.Per follow up via phone on 22 aug 2023, customer experienced a sepsis infection and an urinary tract infection with used.Customer was treated at the emergency room for the sepsis infection and then treated with antibiotics for the urinary tract infection.Customer's husband stated that the customer has multiple sclerosis which caused them body to spasm.
 
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Manufacturer Narrative
The reported event was inconclusive because no sample was returned.The potential root cause for this failure could be due to "dimensions not specified correctly and/or improper materials used assembly collapses under vacuum".It is unknown whether the device had met relevant specifications.The product was used for treatment purposes.It was unknown whether the product had caused the reported failure.The lot number was unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: ¿setup: 1.Connect the canister to wall suction and set to a minimum of 40mmhg continuous suction.Always use the minimum amount of suction necessary.If using the drydoctm vacuum station / purewick¿ urine collection system, connect the canister to the unit and turn the unit on.Please consult the purewick¿ urine collection system user guide for further information.2.Using standard suction tubing, connect the purewick¿ female external catheter to the collection canister.Precautions: not recommended for patients who are: agitated, combative, or uncooperative and might remove the purewicktm female external catheter.Peri-care and placement: 1.Perform perineal care and assess skin integrity (document per hospital protocol).Separate legs, gluteus muscles, and labia.Palpate pubic bone as anatomical marker.2.With soft gauze side facing patient, align distal end of the purewicktm female external catheter at gluteal cleft.Gently tuck soft gauze side between separated gluteus and labia.Ensure that the top of the gauze is aligned with the pubic bone.Slowly place legs back together once the purewicktm female external catheter is positioned.Note: patient can be positioned on back, side lying, frog legged, or lying on back with knees bent and thighs apart (lithotomy position) prior to device placement.Recommendations: ensure the purewicktm female external catheter remains in the correct position after turning the patient.Remove the purewicktm female external catheter prior to ambulation.Properly placing the purewicktm female external catheter snugly between the labia and gluteus holds the purewicktm female external catheter in place for most patients.Mesh underwear maybe useful for securing the purewicktm female external catheter for some patients.Change suction tubing per hospital protocol or at least every thirty (30) days.Do not use barrier cream on the perineum when using the purewicktm female external catheter.Barrier cream may impede suction.Do not use the purewicktm female external catheter with bedpan or any material that does not allow for sufficient airflow".H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Event Description
It was reported that the patient was no longer using because the wick was backing up and not suctioning properly and caused the patient to develop multiple bacterial infections.It was unknown what medical intervention was provided for bacterial infections.Per follow up via phone on 22aug2023, customer experienced a sepsis infection and an urinary tract infection with used.Customer was treated at the emergency room for the sepsis infection and then treated with antibiotics for the urinary tract infection.Customer's husband stated that the customer has multiple sclerosis which caused them body to spasm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PUREWICK FEMALE EXTERNAL CATHETER
Type of Device
PUREWICK FEMALE EXTERNAL CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key17728278
MDR Text Key323161973
Report Number1018233-2023-06703
Device Sequence Number1
Product Code NZU
UDI-Device Identifier00801741189050
UDI-Public(01)00801741189050
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Consumer
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 09/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue NumberPWFX30
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/22/2023
Initial Date FDA Received09/12/2023
Supplement Dates Manufacturer Received09/26/2023
Supplement Dates FDA Received09/27/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexFemale
-
-