MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION WOLVERINE CORONARY CUTTING BALLOON MONORAIL; CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION WOLVERINE CORONARY CUTTING BALLOON MONORAIL; CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING Back to Search Results
Model Number 3851
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/12/2023
Event Type  malfunction  
Event Description
It was reported that the catheter was broken.The target lesion was located in the left anterior descending artery.A 10mmx3.00mm wolverine coronary cutting balloon was selected for use.During the procedure it was noted that the catheter was broken, and it could not enter the patient's body through the guidewire properly.The procedure was completed with another of same device.No complications were reported, and patient was stable post procedure.
 
Manufacturer Narrative
Device evaluated by manufacturer: the device was returned for analysis.A visual examination of the balloon identified no damages, contrast media was inside the balloon.Multiple hypotube kinks were identified along the shaft of the device.No kinks or damages on shaft polymer extrusion.A detailed microscopic examination of the balloon material identified no tears or pinholes in the balloon.All blades were fully bonded on the balloon and did not exhibit any signs of damage.No issues identified during examination of the extrusion shaft.Tip showed no signs of tip damage.A microscopic examination of the proximal and distal markerbands identified no damage.
 
Event Description
It was reported that the catheter was broken.The target lesion was located in the left anterior descending artery.A 10mmx3.00mm wolverine coronary cutting balloon was selected for use.During the procedure it was noted that the catheter was broken, and it could not enter the patient's body through the guidewire properly.The procedure was completed with another of same device.No complications were reported, and patient was stable post procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WOLVERINE CORONARY CUTTING BALLOON MONORAIL
Type of Device
CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC IRELAND LIMITED
ballybrit business park
galway
EI  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key17728733
MDR Text Key323171040
Report Number2124215-2023-47621
Device Sequence Number1
Product Code NWX
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/12/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3851
Device Catalogue Number3851
Device Lot Number0030479483
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/06/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/04/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age69 YR
Patient SexMale
Patient Weight65 KG
-
-