It was reported that the procedure was to treat a mildly calcified, mildly tortuous left internal carotid artery that is 90% stenosed.The vessel was prepped with 4x30mm viatrac balloon to atmospheres for 3 seconds.The 8tx40x136 xact stent knob did not rotate during deployment; therefore, the two surgeons jointly used very large forces to complete the release of the stent.The stent was implanted successfully.There was no adverse patient effects and no clinically significant delay reported.No additional information was provided.
|
The device was returned for analysis.The reported mechanical jam was unable to be confirmed.The reported difficult or delayed activation-stent was unable to be replicated in a testing environment due to the condition of the returned device.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.As there was no damage noted to the device during the inspection prior to use, the investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that interaction with the mildly calcified, mildly tortuous and 90% stenosed anatomy and/or inadvertent mishandling resulted in the noted multiple bends in the shaft preventing the shaft lumens from moving freely; thus resulting in the reported thumbwheel difficulties and the reported difficulty deploying the stent.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|