ABBOTT ENDURITY¿ CORE SR, JPD; PULSE GENERATOR, PERMANENT, IMPLANTABLE Back to Search Results
Model Number PM1152
Device Problems Failure to Interrogate (1332); No Pacing (3268)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/05/2023
Event Type  Injury  
Event Description
During an in clinic follow up, the device was unable to be interrogated using inductive telemetry and a loss of pacing was observed.The device was explanted and replaced to resolve the event.The patient was stable.
 
Manufacturer Narrative
The reported events of no pacing and no inductive telemetry were confirmed.The device was received with no telemetry communication and no output.Visual inspection of the header attachment area detected an anomaly between the epoxy header and titanium case.The device was cut open to enable further testing and the battery was found in normal range.Feedthrough leak test was performed, indicating a device hermeticity breach.This is consistent with feedthrough damage as a result of fluid intrusion between the header and case, and subsequent fluid ingress to internal electronics.Hybrid circuitry was tested, and the results indicated normal current drain.A manufacturing process anomaly may have occurred, which resulted in the header bonding anomaly.The device is included in the assurity and endurity pacemakers header anomaly advisory issued by abbott on 15 march 2021.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDURITY¿ CORE SR, JPD
Type of Device
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17729974
MDR Text Key323192381
Report Number2017865-2023-42845
Device Sequence Number1
Product Code NVZ
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
P880086
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date 11/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2020
Device Model NumberPM1152
Device Lot NumberA000062002
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/05/2023
Initial Date FDA Received09/12/2023
Supplement Dates Manufacturer Received10/27/2023
Supplement Dates FDA Received11/08/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/19/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberFA-Q121-CRM-1
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-