Model Number WB91051W
Device Problem Defective Component (2292)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device was evaluated, and the customer¿s allegation was confirmed.In addition to the e433 code (generator defective), there were no additional findings.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the event was unable to be identified; however, the event most likely occurred due to a faulty generator board.Olympus will continue to monitor field performance for this device.(b)(6).
 
Event Description
It was reported to olympus that the hf unit "esg-400" had an error occur and unit switches itself off.The event was found during a procedure.There was no patient harm associated with the event.The device was evaluated, and it was found that there was an error code e433 (generator board defective).This mdr (medical device report) is being submitted to capture the reportable malfunction found during the device evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF UNIT "ESG-400"
Type of Device
ELECTROSURGICAL GENERATOR ESG-400
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH BERLIN FACILITY
rheinstrasse 8
teltow, brandenburg 14513
GM  14513
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH BERLIN FACILITY
rheinstrasse 8
teltow, brandenburg 14513
GM   14513
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17736097
MDR Text Key323287663
Report Number3003724334-2023-00035
Device Sequence Number1
Product Code GEI
UDI-Device Identifier04042761076838
UDI-Public04042761076838
Combination Product (y/n)N
Reporter Country CodeSF
PMA/PMN Number
K203682
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 09/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/13/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberWB91051W
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received08/24/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/02/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-