MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH TELESCOPE, 10 MM, 0°, HD, QUICK LOCK, AUTOCLAVABLE; THORACOSCOPE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH TELESCOPE, 10 MM, 0°, HD, QUICK LOCK, AUTOCLAVABLE; THORACOSCOPE SYSTEM Back to Search Results
Model Number WA53000A
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/21/2023
Event Type  malfunction  
Event Description
The customer reported that a telescope, 10 mm, 0°, hd, quick lock, autoclavable¿s light guide connector was loose and came off.The event occurred after the procedure.The therapeutic procedure (total laparoscopic hysterectomy) was completed with the same device.There was no patient harm associated with the event.
 
Manufacturer Narrative
The device was returned and evaluated, and the customer¿s allegation (loose light guide connector and comes off) was confirmed.In addition, there was foreign matter adhesion on the internal lens.The investigation is ongoing, and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that the event occurred due to user error, improper handling as well as application of excessive force.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TELESCOPE, 10 MM, 0°, HD, QUICK LOCK, AUTOCLAVABLE
Type of Device
THORACOSCOPE SYSTEM
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17737459
MDR Text Key323291595
Report Number9610773-2023-02557
Device Sequence Number1
Product Code EWY
UDI-Device Identifier04042761052757
UDI-Public04042761052757
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K912362
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 12/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/13/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberWA53000A
Device Lot Number541588
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/21/2023
Was the Report Sent to FDA? No
Date Manufacturer Received12/21/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/17/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-