Model Number N/A
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Pain (1994); Scar Tissue (2060)
Event Date 09/04/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4).D10: 51-113140 item name tprlc 133 type1 bm so 14.0 lot # 6057581, 010000707 item name g7 bonemaster ltd acet shl 60g lot # 6332149.G2: foreign: australia multiple mdr reports were filed for this event, please see associated reports: 0001825034 - 2023 - 02162, 0001825034 - 2023 - 02161.The device will not be returned for analysis as it remains implanted; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch will be submitted h3 other text : device remains implanted.
 
Event Description
It was reported that the patient underwent a revision procedure 7 months post implantation due to pain, possible poly disimpaction and instability.During the surgery, the poly was found intact and extensive scar tissue was debrided.Only the head was revised, and the procedure was completed without complication.There is no additional information available at the time of this report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Complaint confirmed based on evaluation of the provided medical records.Review of the device history record(s) identified no deviations or anomalies during manufacturing.Medical records/radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: x-ray review: there is no abnormal radiolucency or evidence of implant loosening.There is malalignment with subluxation/dislocation as noted.Records review: unstable, extensive scar tissue- debrided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
There is no update to the prior event description provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 10 DEG E1 LINER 36MM G
Type of Device
PROSTHETIC, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key17739786
MDR Text Key323324076
Report Number0001825034-2023-02160
Device Sequence Number1
Product Code PBI
UDI-Device Identifier00880304526785
UDI-Public(01)00880304526785(17)220203(10)3959772
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date02/03/2022
Device Model NumberN/A
Device Catalogue Number010000898
Device Lot Number3959772
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/21/2023
Initial Date FDA Received09/13/2023
Supplement Dates Manufacturer Received10/30/2023
Supplement Dates FDA Received11/02/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/01/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient SexMale
Patient Weight120 KG
-
-