Catalog Number 03666794001 |
Device Problem
Image Display Error/Artifact (1304)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/18/2023 |
Event Type
malfunction
|
Event Description
|
There was an allegation of a meter display issue on a coaguchek xs meter, serial number (b)(6).The customer alleged that the meter display is faded even after changing the batteries and that the first number in the meter memory and the countdown are missing segments.The customer performed a display check and they stated they see a flashing strip but not much else.There was no allegation of misinterpreted results due to the display issue.
|
|
Manufacturer Narrative
|
The product has been requested for investigation.The investigation is ongoing.
|
|
Manufacturer Narrative
|
Sections d9 and h3 were updated.The meter was received for investigation.The investigation found that the meter circuit board was contaminated and damaged by penetrative liquid contamination due to user mishandling.No product problem was identified.
|
|
Search Alerts/Recalls
|