Model Number MR290V
Device Problems Crack (1135); Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/06/2023
Event Type  malfunction  
Event Description
A healthcare facility in china reported that a mr290v vented autofeed humidification chamber was found leaking water and cracked before patient use.There was no patient involvement.
 
Manufacturer Narrative
(b)(4).The complaint mr290v vented autofeed humidification chamber is currently en route to fisher & paykel healthcare (f&p) for evaluation.We will provide a follow up report upon completion of investigation.
 
Manufacturer Narrative
(b)(4).Method: the complaint mr290v vented autofeed humidification chamber was not returned to fisher & paykel healthcare for evaluation.Our investigation is therefore based on the information provided by the customer and our knowledge of the product.Results: the customer reported that a mr290v vented autofeed humidification chamber was found leaking water.Further information was provided by the customer that the water was leaking at the connection of water feedset tube and was not leaking in large quantity.Conclusion: without the complaint device, we are unable to determine the cause of the reported event.The mr290v chambers are designed and tested to conform to iso 5367 breathing tubes intended for use with anaesthetic apparatus and ventilators.Every mr290v chamber is pressure tested following the manufacturing process to check for any leaks present in the chamber dome due to cracks and other causes.Any chamber which fails this test is rejected.In addition, the pressure test is followed by a visual inspection of each chamber.No cracks in the chamber dome are acceptable.Any chamber that fails this inspection is rejected.The subject mr290v chamber would have met the required specification at the time of production.Our user instructions that accompany the mr290v vented autofeed humidification chamber state the following: - "do not soak, wash, sterilize, or reuse this product.Avoid contact with chemicals, cleaning agents, or hand sanitizers." - "ensure appropriate ventilator or flow source alarms are set before connecting breathing set to patient." - "perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient." - "use of the mr290 above the maximum operating pressure may lead to cracking, water leakage and, on rare occasions, could lead to a loss of ventilation pressure.".
 
Event Description
A healthcare facility in china reported that a mr290v vented autofeed humidification chamber was found leaking water and cracked before patient use.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENTED AUTOFEED HUMIDIFICATION CHAMBER
Type of Device
AUTOFEED CHAMBER
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
omid taheri
17400 laguna canyon road
suite 300
irvine, CA 92618
8007923912
MDR Report Key17741913
MDR Text Key323660542
Report Number9611451-2023-00859
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K934140
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMR290V
Device Catalogue NumberMR290V
Device Lot Number2102221910
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/10/2023
Initial Date FDA Received09/13/2023
Supplement Dates Manufacturer Received10/20/2023
Supplement Dates FDA Received10/24/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
F&P RT380 ADULT EVAQUA2 BREATHING CIRCUIT; F&P RT380 ADULT EVAQUA2 BREATHING CIRCUIT
-
-