Catalog Number SL-2000M2095L
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
As reported by the user facility: venus pod rupture into internal transducer of machine.No injury reported.
 
Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).No samples or photos were submitted to the manufacturer for evaluation.A device history record (dhr) review was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.Reviewing the history records during an internal review, it was determined that the root cause of the defect is the machine; due to a high pressure of return that can be stressed and generate a perforation in the pod membrane.Based on the history records, the reported defect is confirmed.An approved project is in place to further address issues with pod internal breakthrough.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMODIALYSIS BLOODLINES
Type of Device
SET, TUBING, BLOOD, WITH A
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
824 12th avenue
bethelehem PA 18018
Manufacturer (Section G)
B. BRAUN MEDICAL INC.
824 12th avenue
bethelehem PA 18018
Manufacturer Contact
jonathan severino
901 marcon blvd.
allentown, PA 18109
4847197287
MDR Report Key17743300
MDR Text Key323910028
Report Number2521402-2023-00229
Device Sequence Number1
Product Code FJK
UDI-Device Identifier04046955348909
UDI-Public(01)04046955348909
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K080807
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberSL-2000M2095L
Device Lot NumberA2200647
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/21/2023
Initial Date FDA Received09/14/2023
Supplement Dates Manufacturer Received08/21/2023
Supplement Dates FDA Received03/19/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured08/07/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-