• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NALU MEDICAL INC NALU NEUROSTIMULATION SYSTEM; SPINAL CORD STIMULATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NALU MEDICAL INC NALU NEUROSTIMULATION SYSTEM; SPINAL CORD STIMULATOR Back to Search Results
Model Number 71005
Device Problem Migration (4003)
Patient Problem Inadequate Pain Relief (2388)
Event Date 08/15/2023
Event Type  Injury  
Event Description
Patient was implanted with the nalu spinal cord stimulator system on (b)(6) 2023.Patient had successful activation but then subsequently reported inadequate pain relief due to difficulty maintaining communication between the implantable pulse generator (ipg) and the external therapy discs.Communication issues were determined to be due to migration of the ipg within the pocket.Surgical revision procedure was performed on (b)(6) 2023 to remove the ipg from the existing pocket and place it into a new pocket.All existing components were utilized, and no components were replaced during the procedure.
 
Manufacturer Narrative
Migration is a known inherent risk of implantable neuromodulation systems.There are no allegations of any malfunction of the system or its components and all components remain implanted at the time of this reporting.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NALU NEUROSTIMULATION SYSTEM
Type of Device
SPINAL CORD STIMULATOR
Manufacturer (Section D)
NALU MEDICAL INC
2320 faraday avenue
suite 100
calsbad CA 92008 2377
Manufacturer (Section G)
NALU MEDICAL, INC
2320 faraday ave
suite 100
carlsbad CA 92008 7241
Manufacturer Contact
terry villarba
2320 faraday avenue
suite 100
carlsbad, CA 92008-7241
7604482360
MDR Report Key17743595
MDR Text Key323375167
Report Number3015425075-2023-00217
Device Sequence Number1
Product Code GZB
UDI-Device Identifier00812537036137
UDI-Public01008125370361371123020817260208
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K183579
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/14/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number71005
Device Catalogue Number71005
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/15/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/08/2023
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-