LIVANOVA DEUTSCHLAND GMBH S5 CONSOLE FOR 4 PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 48-40-00
Device Problem Pumping Stopped (1503)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/18/2023
Event Type  malfunction  
Event Description
Livanova deutschland has received a report that the outflow pump magnet decoupled from the circuit and did not flow blood back to the patient during a procedure.There is no report of any patient injury.Device was took out of service and used back up.
 
Manufacturer Narrative
A1-a5: patient information was not provided.H10: livanova deutschland manufactures the s5 console.The incident occurred in united states.A livanova field service engineer was dispatched to the customer site.After testing the scp unit, the device is working as intended.No issues were found at this time.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
See initial report.
 
Manufacturer Narrative
H10: complaints database analysis revealed that no similar event on this device occurred since its installation in 2014.Based on the collected data and taking into account the technical service intervention, a hardware failure of the centrifugal pump system can be excluded.It cannot be ruled out that the root cause of the flow reduction could be due to a specific condition related to procedure, patient or a disposable failure.This report was due on november 14, 2023; however, due to a network disruption at livanova, the ability to submit mdrs was lost on november 19, 2023, before this report was ready to submit.The report was prepared and submitted immediately following restoration of the livanova systems.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 CONSOLE FOR 4 PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND GMBH
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80939
GM   80939
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key17744544
MDR Text Key323388882
Report Number9611109-2023-00438
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K210130
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/14/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number48-40-00
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received10/25/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/30/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-