Model Number 0998-00-3023-53 |
Device Problem
Premature Discharge of Battery (1057)
|
Patient Problems
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/07/2023 |
Event Type
malfunction
|
Event Description
|
It was reported by customer that, the cs300 intra-aortic balloon pump (iabp) units battery lasts for about five minutes.There was no patient involvement reported.
|
|
Manufacturer Narrative
|
A supplemental report will be submitted upon completion of our investigation.
|
|
Event Description
|
It was reported that while walking the patient the cs300 intra-aortic balloon pump (iabp) got a low battery alarm.Units battery lasts for about five minutes.They took patient back to room and plugged device into power.
|
|
Manufacturer Narrative
|
Updated data : b4,g3,g6,h2,h10,h11.Corrected b5,b6,b7,d10,h6(impact,clinical).
|
|
Event Description
|
It was reported that while walking the patient the cs300 intra-aortic balloon pump (iabp) got a low battery alarm.Units battery lasts for about five minutes.They took patient back to room and plugged device into power.There was no patient harm.
|
|
Manufacturer Narrative
|
A getinge field service engineer (fse) evaluated the intra-aortic balloon pump and replaced the batteries and performed full functional and safety tests.All tests passed.The fse returned the device to the customer and cleared for clinical use.
|
|
Event Description
|
N/a.
|
|