• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN MFG DC BOULDER UNKNOWN COOL TIP ELECRTRODE; Electrosurgical, cutting & coagulation & accessories

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MFG DC BOULDER UNKNOWN COOL TIP ELECRTRODE; Electrosurgical, cutting & coagulation & accessories Back to Search Results
Model Number UNKNOWN COOL TIP ELECRTRODE
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Hemorrhage/Bleeding (1888); Hemothorax (1896); Pleural Effusion (2010); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 02/12/2023
Event Type  Injury  
Manufacturer Narrative
D10 concomitant product: unknown cool-ti, unknown cool-tip rf ablation generator, (serial #unknown) reference: (b)(6), outcomes of laparoscopic radiofrequency ablation versus percutaneous radiofrequency ablation for hepatocellular carcinoma; (b)(6); surgical endoscopy (2023) 37:5176¿5189, https://doi.Org/10.1007/s00464-023-09956-1.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the literature, a retrospective study was conducted between (b)(6) 2005 and (b)(6) 2020 to compare outcomes of laparoscopic radiofrequency ablation (lrfa) and percutaneous radiofrequency ablation methods (prfa) in patients with hepatocellular carcinoma.A total of 381 patients underwent prfa or lrfa using a cool-tip or a non-medtronic electrode.The following complications were reported in both rfa groups: pleural effusion, hemothorax, perihepatic bleeding and bowel injury.The author did not mention any interventions used to treat these complications.The article did not specify which electrode was associated with the reported complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN COOL TIP ELECRTRODE
Type of Device
Electrosurgical, cutting & coagulation & accessories
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
Manufacturer (Section G)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key17746488
MDR Text Key323418340
Report Number1717344-2023-01155
Device Sequence Number1
Product Code GEI
Combination Product (y/n)N
Reporter Country CodeKS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Literature,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 09/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberUNKNOWN COOL TIP ELECRTRODE
Device Catalogue NumberUNKNOWN COOL TIP ELECRTRODE
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 08/25/2023
Initial Date FDA Received09/14/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SEE H10.
Patient Outcome(s) Other;
-
-