A visual assessment of the returned system inserter/compressor showed an instrument with repeated use, as identified by surface scratches.The compressor handle ratchet and one distal tip of the instrument were fractured from the body of the system inserter/compressor.A functionality assessment was not performed due to the damaged condition of the returned system inserter/compressor, which was removed from distributable inventory.A dhr review was performed for the instrument lot# and there were no manufacturing anomalies identified.The instrument lot met all required specifications prior to being released to distributable inventory.This lot has been available for distribution since 2/10/2021.It was reported that the system inserter/compressor was impacted by the surgeon to place the implant more anterior.It is unknown why the surgeon elected to use a mallet to adjust the implant location more anterior.The surgical system does not contain a mallet, and the system inserter/compressor is not designed to be impacted for implant placement.The root cause of this complaint is impacts to the system inserter/compressor which resulted in the observed instrument malfunction.There have not been any other complaints of similar nature in the past 12 months.The manufacturer will continue to monitor this instrument for complaints from the field.
|