A facility reported that during "c3-7 removal of instrumentation, c3-t2 posterior decompression with durotomy repair and neuro monitoring" on a patient, the mayfield modified skull clamp (a1059) slipped.The device was in contact with the patient,but there was no injury.There was a 10 minute surgery delay as the surgeon had to reposition the patient's head after starting the procedure.
|
the mayfield modified skull clamp (a1059) was returned for evaluation: failure analysis - investigation of the returned unit showed that the lock had rotational and lateral movement and a residue buildup was present.The starburst teeth were worn down, so the base casting was replaced.The index knob was revised to the current edition, and new components were added to replace worn internal parts along with general cleaning and maintenance.Root cause - the complaint could not be confirmed regarding the slippage, but the complaint on the starburst teeth is confirmed.Probable root cause is routine use and wear.Additionally, improper or suboptimal placement of the skull clamp can lead to movement of slippage of the patient.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.At present, we consider this complaint to be closed.
|