• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 SITE-RITE 8 ULTRASOUND SYSTSEM CONSOLE (REMANUFACTURED); SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 SITE-RITE 8 ULTRASOUND SYSTSEM CONSOLE (REMANUFACTURED); SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC Back to Search Results
Model Number N/A
Device Problems No Display/Image (1183); No Apparent Adverse Event (3189)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/06/2023
Event Type  malfunction  
Event Description
It was reported the bottom of the screen is practically black and the top of the screen is difficult to see as well.
 
Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The date of event was not provided by the complainant/reporter, the date reflected in this report is the date bd became aware of the event.The device has not been received by the manufacturer for evaluation.
 
Event Description
It was reported the bottom of the screen is practically black and the top of the screen is difficult to see as well.
 
Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The date of event was not provided by the complainant/reporter, the date reflected in this report is the date bd became aware of the event.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling and applicable manufacture records.Based on a review of the information, the following was concluded: the device was returned to service facility for evaluation.During evaluation, the reported issue of the bottom of the screen is very dark, top of screen is blurry was unconfirmed.The reported issue could not be confirmed.There is no root cause as the issue could not be reproduced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SITE-RITE 8 ULTRASOUND SYSTSEM CONSOLE (REMANUFACTURED)
Type of Device
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
DYMAX CORP. -2523003
110 marshall drive
warrendale PA 15086
Manufacturer Contact
becky garcia
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key17752953
MDR Text Key323469094
Report Number3006260740-2023-04097
Device Sequence Number1
Product Code IYO
UDI-Device Identifier00801741138287
UDI-Public(01)00801741138287
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K182281
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number9770600Y
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/06/2023
Initial Date FDA Received09/14/2023
Supplement Dates Manufacturer Received10/24/2023
Supplement Dates FDA Received10/26/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-