Model Number 71992-01 |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Itching Sensation (1943); Skin Inflammation/ Irritation (4545)
|
Event Date 08/30/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
The reported product is not expected to be returned as the device was reportedly discarded.A follow-up report will be submitted if product is returned or additional information is obtained.The device manufacturing date is unknown.The date entered in section h4 is the date abbott diabetes care became aware of the event.The date the incident occurred is unknown.The date entered in section b3 is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
A customer experienced a skin reaction while wearing an adc device.Customer reported skin redness and itching and after 10 days of use customer removed the sensor and there was a ¿strong allergy¿ and a bad smell.Customer had contact with a healthcare professional and was provided antibiotic clavox and fusibact gel for treatment.There was no report of death or permanent impairment associated with this event.
|
|
Manufacturer Narrative
|
Repeated attempts by adc to retrieve the product were unsuccessful and/or the customer discarded the product.The most probable root causes associated with this failure mode are disconnected, faulty or damaged components, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.At this time product has not yet been returned.An extended investigation has been performed for the reported complaint.There was no indication that the product did not meet specification.The reported complaint is related to skin irritation or an allergic reaction to the patch adhesive of the freestyle libre sensor.The dhrs (device history review) for the fs libre sensor kit were reviewed and the dhrs showed the sensor kit passed all tests prior to release.Dose audit reports were reviewed and demonstrates the continued effectiveness of the established sterilization process for libre sensor kits.Environmental monitoring reports were reviewed, including bioburden and endotoxin testing, and demonstrated that all monitoring processes continue to meet adc minimum requirements for product quality.In the event that unanticipated product is received, a physical investigation will be performed per adc's established processes and procedures and a follow-up report will be submitted upon completion of investigation.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
A customer experienced a skin reaction while wearing an adc device.Customer reported skin redness and itching and after 10 days of use customer removed the sensor and there was a ¿strong allergy¿ and a bad smell.Customer had contact with a healthcare professional and was provided antibiotic clavox and fusibact gel for treatment.There was no report of death or permanent impairment associated with this event.
|
|
Manufacturer Narrative
|
Repeated attempts by adc to retrieve the product were unsuccessful and/or the customer discarded the product.The most probable root causes associated with this incident are the adhesive, including the adhesive irritating the user¿s skin, or misuse, including improper site selection and repeatedly using the same application site to place the sensor.These conditions are mitigated through the freestyle product labelling.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.At this time product has not yet been returned.Extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.The reported complaint is related to skin irritation or an allergic reaction to the patch adhesive of the freestyle libre sensor.Dose audit reports were reviewed and demonstrates the continued effectiveness of the established sterilization process for libre sensor kits.Environmental monitoring reports were reviewed, including bioburden and endotoxin testing, and demonstrated that all monitoring processes continue to meet adc minimum requirements for product quality.The dhrs (device history review) for the libre sensor and libre sensor kits were reviewed and the dhrs showed the libre sensor and libre sensor kits passed all tests prior to release.In the event that unanticipated product is received, a physical investigation will be performed per adc's established processes and procedures and a follow-up report will be submitted upon completion of investigation.This serves as a correction report.Section h10 (addtl mfg narrative) was incorrectly documented in the previous report.Correction has been made.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
A customer experienced a skin reaction while wearing an adc device.Customer reported skin redness and itching and after 10 days of use customer removed the sensor and there was a ¿strong allergy¿ and a bad smell.Customer had contact with a healthcare professional and was provided antibiotic clavox and fusibact gel for treatment.There was no report of death or permanent impairment associated with this event.
|
|
Search Alerts/Recalls
|